Page 84 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02
Parameters
P924 Volume Measurement
The calculated vessel capacity in Max Volume (P051) or % of Max Volume.
Press ENTER as required to display in Max. Volume units.
P927 Distance Percent
The distance between the surface and the transducer face.
Use P923 unless the distance information is required in percent (press ENTER ).
Master Reset (P999)
This feature resets all parameters to original values.
Use this feature prior to initial programming if arbitrary Parameter Values were used
during a bench test, or after upgrading the software. Following a Master Reset, complete
reprogramming is required.
To perform a Master Reset, access P999 and press CLEAR . C.ALL displays until
the reset is complete.
Values Range: 0.000 to 9999
CAUTION: Be careful when using this feature. All data will be reset.
For convenience, be sure to record the values you want to re-enter.