Page 204 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01
mmmmm
Parameters
P904 Keypad
Press ENTER , then press each keypad key in the following sequence:
As each key is pressed, the associated keypad number is displayed. On successful test
completion, PASS is displayed. FAIL is displayed if a key is pressed out of sequence or the
programmer keypad malfunctions.
P905 Transmit Pulse
This feature may be used to monitor the transmit pulse with an oscilloscope connected to
the transducer terminals.
Press ENTER to supply repeated transmit pulses, at the frequency entered, to the
transducer.
P906 RS-232 Port
Press ENTER
to test the RS-232 port on the RJ-11.
An external device must be connected to the RS-232 port for this test. On successful
completion, PAS S is displayed, otherwise it is FAIL.
P908 Scanner
Press ENTER to cycle the scanner relay while firing the transmitter.
Use this parameter to ensure that both transducers are being stimulated.
Primary Index Global
Values Range: 42kHz to 46kHz (view only)
Altered By P004 Transducer
P
P